SKC 224-PCXR8 Air Sampling Pump

Rent for:$25-Day$95-Week$285-month
 
SKC Logo
SKC_PCXR8_Manual.pdf
File Size: 3521 kb
File Type: pdf
Download File

SKC_PCXR8_Series_DataSheet.pdf
File Size: 151 kb
File Type: pdf
Download File

SKC 224-PCXR8
 

224-PCXR8 Air Sampling Pump Specifications

Operating Range

1000 to 5000 ml/min (UL Listed model)
1000 to 3000 ml/min (MSHA-approved model)
(5 to 500 ml/min requires the 224-26 Series Adjustable Low Flow Holder)

Weight
34 oz (964 gm)
Dimensions
5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.8 cm)
Compensation Range
1000 to 2500 ml/min at 40 inches water back pressure
3000 ml/min at 35 inches water back pressure
4000 ml/min at 20 inches water back pressure
5000 ml/min at 10 inches water back pressure
Flow Control
Holds constant flow to ± 5% of set-point
Run Time: Battery
8 Hours minimum at 4000 ml/min and 20 inches water back pressure; dependent on media used.
Run Time: Mains Adapter
9999 minutes (6.8 days). Pump will shut off as run time cannot exceed timer display (see Timer Display Range below)
Timer Display Range:
1 to 9999 minutes (6.8 days). Pump will shut off at 9999 minutes. To reset, restart the pump.
Flow Indicator
Built-in rotameter with 250-ml division; scale marked at 1, 2, 3, 4, and 5 L/min
Charging Time
(varies with capacity and level of charge)
6 to 8.5 hrs with PowerFlex charger
Power Supply
6.0 V plug-in NiCad battery pack, rechargeable, 2.0-Ah capacity
Intrinsic Safety

UL Listed for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and Class III, Temperature Code T3C
ATEX-approved model is available;

Operating Temperature
32 to 113 F (0 to 45 C)
Storage Temperature
-40 to 113 F (-40 to 45 C)
Charging Temperature
50 to 113 F (10 to 45 C)
Operating Humidity
0 to 95% Relative
Flow and Low Battery Fault
Fault shutdown on excessive back pressure or low battery with LCD indicator and time display retention
Battery Test
LCD indicator verifies battery condition prior to sampling
Time Display
LCD indicator displays sampler run time in minutes
Delay On
Allows user to select minutes to delay start of test up to 9999 minutes (6.8 days)
Timed Shutdown
Allows user to select minutes of operation before automatic shutdown. Timed shutdown maximum is 9999 minutes (6.8 days)
Timing Accuracy
± 0.05% (± 45 sec/day)
Intermittent Sampling
Programmable to allow user to extend short-term samples over an extended period of time to meet time weighted average (TWA) requirements with a reduced number of samples. Elapsed time maximum is 9999 minutes (6.8 days), at which time the sample pump shuts down.
Sampling Pause
Allows user to temporarily halt sampling without loss of timing data. Restart does not require resetting time.
Multiple-tube Sampling
Built-in constant pressure regulator allows user to take up to four simultaneous samples with rates up to 500 ml/min each (total combined flow 1350 ml/min maximum) using the optional adjustable low flow holder accessory
RFI/EMI Shielding
Complies with requirements of EN 55022, FCC Part 15 Class B, EN 50082-1
Frequency range of the radiated susceptibility test was 27 to 1000 MHz
CE marked
Additional Specifications
Built-in fluid/particulate trap
External see-through filter housing
Tamper-resistant control cover
Tamper-resistant compensation adjustments
Built-in accessory mount
Stainless steel belt clip
External exhaust port
All non-plastic parts made of corrosion-resistant materials
RFI and EMI-shielded case