LaMotte 2020e and 2020i Turbidity Meters
The multi-detector optical configuration assures long term stability and minimizes stray light and color interferences. All readings are determined by the process of signal averaging over a 5 second period. This minimizes fluctuations in readings attributed to large particles and results in rapid, highly repeatable measurements. Ideally suited for both low-level drinking water applications as well as monitoring high turbidity in the field.
|
To Order Call 1-877-386-2480
...request a quote online |
- Patent pending optical design features focusing optics for low level precision and accuracy.
- Six user selected languages – English, Spanish, French, Japanese, Italian, and Portuguese.
- MSP430 Microcontroller used is the most advanced controller on the market for hand held applications.
- Advanced calibration algorithms.
- Tube positioning ring limits vial variability.
- Easy menu driven operation and large LCD display.
- 4000 point data log, stored results can be viewed directly on instrument or downloaded to a computer via RS232 cable.
|
To Order Call 1-877-386-2480
...request a quote online | 2020e version meets USEPA design criteria as specified by USEPA method 180.1.
2020i version meets design criteria for quantitative methods of turbidity using optical turbidimeters as specified by ISO 7027. Kits are supplied with 0, 1, and 10 NTU standard, sample bottle, four sample tubes, and an extra battery. AC power adapter is optional. |
Turbidity Specifications: |
|
|
Unit of Measure: |
NTU, FNU, FAU, ASBC, EBC |
|
Range: |
0-4000 |
|
Resolution: |
0.01 NTU/FNU 0.00-10.99, 0.1 NTU/FNU 11.00-109.9, |
|
Accuracy: |
±2% 0-100 NTU, ±3% above 100 NTU |
|
Detection Limit: |
0.05 NTU/FNU |
|
Range Selection: |
Automatic |
|
Reproducibility: |
0.02 NTU/FNU, 0.5 FAU |
|
Light Source: |
860nm LED (2020i ISO) |
Meter Features: |
|
|
Signal Averaging: |
Disabled, 2, 5, 10 |
|
Battery: |
9V |
|
AC Power: |
Optional |
|
Data Logging: |
4000 points |
|
Auto Shut-Off: |
Disabled, 5, 10, 30 |
|
Optional Software: |
SmartLink 2 |
|
Languages: |
English, French, Spanish, Japanese, Italian, Portuguese |
|
Response Time: |
<5 Seconds |
Item Description |
Order Code |
|
Model 2020e Kit |
1979-EPA |
|
Model 2020i Kit |
1979-ISO |
|
0 NTU Standard (ISO and EPA), 60 mL |
1480 |
|
1 NTU Standard (ISO), 60 mL |
1481 |
|
10 NTU Standard (ISO), 60 mL |
1482 |
|
100 NTU Standard (ISO), 60 mL |
1483 |
| 1 NTU Standard (EPA), 60 mL | 1484 |
| 10 NTU Standard (EPA), 60 mL | 1485 |
| 100 NTU Standard (EPA), 60 mL | 1486 |
| Formazin Standard Solution, 4000 NTU, 60 mL | 6195-H |
| AC Power Adapter (variable 100-240 V AC) | 1754 |
| Vial Ring (2-pk) | 0641 |
| Six-pack of vials | 0475 |
| RS232 Cable | 1772 |
| 2020e&i Meter Manual FREE in downloadable PDF format (Just Click Here) | 1979-MN.PDF |
| 2020e&i Meter CE Compliance Certifate in downloadable PDF format | 2020 CE Certificate |
| AMCO Turbidity Standards, ISO 7027 Certificate of Compliance | ISO7027.PDF |