Foxboro / Thermo Environmental TVA 1000 - Portable Flame Ionization Detector
| Rent for: | $80-Day | $320-Week | $1,000-month |
Detects organic vapors from 1 to 50,000ppm. It has an internal, rechargeable battery and internal sampling pump. The instrument require 99.9% hydrogen to fuel the internal flame which is included with the rental (special shipping required). The unit is calibrated with methane prior to leaving our office. (Intrinsically safe for Class I, Division 2, Groups A, B, C, D)
Click for Larger Image
Manufacturers Website
Manufacturers Website
Accuracy:
Repeatablility:
Analog Output:
Dynamic Range:
Linear Range:
Minimum Detectable Level:
Repeatablility:
Analog Output:
Dynamic Range:
Linear Range:
Minimum Detectable Level:
±25% of reading or ±2.5 ppm, whichever is greater, from 1.0 to 10,000 ppm. Accuracy listed is achieved using methane with a 1-point calibration in the range from 100 to 500 ppm (including drift) at the temperature and humidity of the calibration.
±2% at 100 ppm of methane
Two analog output signals, 0 to 2 dc, proportional to the count output from each detector.
1.0 to 50,000 ppm of methane
1.0 to 10,000 ppm of methane
The minimum detectable level is defined as two times the peak-to-peak noise.
300ppb of hexane
±2% at 100 ppm of methane
Two analog output signals, 0 to 2 dc, proportional to the count output from each detector.
1.0 to 50,000 ppm of methane
1.0 to 10,000 ppm of methane
The minimum detectable level is defined as two times the peak-to-peak noise.
300ppb of hexane
Dynamic Range (PID): 0.5-2,000 ppm isobutylene
Dynamic Range (FID): 0.5-50,000 ppm methane
Linear Range (PID): 0.5-500 ppm isobutylene
Linear Range (FID): 0.5-10,000 ppm methane
Detection Limit (PID): 100 ppb benzene
Detection Limit (FID): 300 ppb hexane
Accuracy (PID): ± 1%
Accuracy (FID): ± 2%
Datalogging: 800-18,000 points mode specific
Weight: 13 pounds (5.8 kg)
Battery Type: Rechargeable Ni-Cad battery
Charging Time: 16 hrs first charge; 2hr charge/1hr use there after
Operating Time: Up to 8 hours S
ampling Rate: 1,000 cc/minute
Listings: FM (Class 1, Div. 1, Groups A,B,C&D Hazardous Location, Temp. Class T4) CENELEC (Div. 1, Zones I and II Group IIC, Hazardous Location, Temp. Class T4)*
Dynamic Range (FID): 0.5-50,000 ppm methane
Linear Range (PID): 0.5-500 ppm isobutylene
Linear Range (FID): 0.5-10,000 ppm methane
Detection Limit (PID): 100 ppb benzene
Detection Limit (FID): 300 ppb hexane
Accuracy (PID): ± 1%
Accuracy (FID): ± 2%
Datalogging: 800-18,000 points mode specific
Weight: 13 pounds (5.8 kg)
Battery Type: Rechargeable Ni-Cad battery
Charging Time: 16 hrs first charge; 2hr charge/1hr use there after
Operating Time: Up to 8 hours S
ampling Rate: 1,000 cc/minute
Listings: FM (Class 1, Div. 1, Groups A,B,C&D Hazardous Location, Temp. Class T4) CENELEC (Div. 1, Zones I and II Group IIC, Hazardous Location, Temp. Class T4)*
| tva_1000_response_factors.pdf |
| tva_1000b_quickstart_guide.pdf |
| tva1000b_manual.pdf |
- Improved Efficiency
- simultaneous photoionization and flame ionization
- on-board datalogging
- multipoint calibration
- multiple calibration curves
- Optional Enhanced Probe
- 16 x 20 character readout
- flame re-ignite
- menu-driven probe operation
- routing functions
- Simple to Use
- menu-driven operation
- easy-to-read LCD displays
- large keypad
- datalogging thumb-switch on hand-held sample probe
- automatic autoranging
- diagnostics
- Optional Enhanced Probe
- 16 x 20 character readout
- flame re-ignite
- menu-driven probe operation
- routing functions